| Место происхождения: | КИТАЙ |
| Фирменное наименование: | AI |
| Сертификация: | ISO |
| Model Number: | AIC-4780 |
| Количество мин заказа: | 1 набор |
|---|---|
| Цена: | Договоренно |
| Packaging Details: | Total Sediment Measuring Instrument packed in hardboard |
| Время доставки: | 20 дней |
| Условия оплаты: | L/C, T/T, Western Union |
| Поставка способности: | 20 комплектов в месяц |
| Test standard: | ASTM D4780,ISO 10307-2,SH/T0701,SH/T0702 | Метод испытаний: | Стареть и горячая фильтрация |
|---|---|---|---|
| Количество образца: | 25 г | Измеренный диапазон: | 0,5% (m/m) |
| Абсолютное давление вакуума: | 40Kpa | Работая untis: | Два |
| Keyward: | Total Sediment Measuring Instrument | Гарантия: | 1 год |
ASTM D4780 Total Sediment Measuring Instrument
![]()
Introduction
This Total Sediment Measuring Instrument is used to determine the total sediment in residual fuel oil with a viscosity of less than 55 mm2/s at 100°C and distillate fuel oil blended with residual components. The maximum total sediment of residual fuel oil is 0.5% (m/m), and the maximum total sediment of distillate fuel oil blended with residual components is 0.40% (m/m)..
Structure
1. Aging bath: electric heating oil bath that can keep 100ºC±0.5ºC.
2. Filter element: made of brass, with copper steam coil attached. The filter element is set up above the vacuum flask.
3. Thermometer: sub-immersion type, the immersion line is 50mm~80mm, and the maximum length is 220mm. The thermometer should be able to record temperatures ranging from 95°C to 105°C. The maximum graduation value is 0.5ºC.
4. Oven: The control temperature is 110ºC±1ºC. It can evaporate the solvent, but there is no danger of fire and explosion.
5. Stirring rod: made of glass or polytetrafluoroethylene, about 150mm in length and 3mm in diameter.
6. Glass beaker: 30ml, either with low mouth or cone. 50ml, narrow neck.
7. Weighing bottle: with frosted glass stopper, number, diameter 80mm, height 40mm.
8. Vacuum source: It can provide the specified absolute pressure of 40kPa±2kPa (61.3kPa vacuum degree).
9. Vacuum meter: It can record the specified absolute pressure or vacuum degree.
10. Glass fiber filter paper: nominal aperture 0.0016mm, diameter 47mm.
11. Cooling container: desiccator or other tightly capped container, used for cooling filter paper before weighing. No desiccant is used.
12. Syringe or washing bottle with scale: the minimum capacity is 25ml, and the graduation value is 0.5ml.
13. Tweezers: flat type.
14. Balance: Sensitivity 0.1mg, used for weighing filter paper. Sensitivity 10mg, used for weighing samples.
II.Technical specifications
| 1 | Sample quantity | 25g |
| 2 | Volume of waste flask | 500ml |
| 3 | Measuring range | 0.5%(m/m) |
| 4 | Test units | Two |
| 5 | Absolute vacuum pressure | 40Kpa |
| 6 | Applicable standards | ASTM D4780,ISO 10307-2,SH/T0701,SH/T0702 |
| 7 | Power supply | AC220V 50HZ |